ASIA unversity:Item 310904400/108283
English  |  正體中文  |  简体中文  |  全文筆數/總筆數 : 94286/110023 (86%)
造訪人次 : 21659196      線上人數 : 507
RC Version 6.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜尋範圍 查詢小技巧:
  • 您可在西文檢索詞彙前後加上"雙引號",以獲取較精準的檢索結果
  • 若欲以作者姓名搜尋,建議至進階搜尋限定作者欄位,可獲得較完整資料
  • 進階搜尋


    請使用永久網址來引用或連結此文件: http://asiair.asia.edu.tw/ir/handle/310904400/108283


    題名: Impacts of 3D bone-to-implant contact and implant diameter on primary implant stability of dental implant
    作者: Huang, 許瑞廷;Jui-Ting Hsu;Yen-Wen Shen;Yen-Wen Shen;Chih-Wei Kuo;Chih-Wei Kuo;Ruei-Teng Wa;Ruei-Teng Wang;Lih-Jyh Fuh;Lih-Jyh Fuh;黃恆立;Heng Li
    貢獻者: 生物資訊與醫學工程學系
    日期: 2017
    上傳時間: 2017-11-03 06:06:46 (UTC+0)
    摘要: Background/Purpose

    This study investigated the effects of three three-dimensional (3D) bone-to-implant contact (BIC) parameters—potential BIC area (pBICA), BIC area (BICA), and 3D BIC percentage (3D BIC%; defined as BICA divided by pBICA)—in relation to the implant diameter on primary implant stability, as well as their correlations were also evaluated.
    Methods

    Dental implants with diameters of 3.75, 4, 5, and 6 mm and artificial bone specimens were scanned by microcomputed tomography to construct 3D models for calculating pBICA, BICA, and 3D BIC%. Indexes of the primary implant stability including the insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ) were measured after implants with various diameters were placed into bone specimens. The Kruskal–Wallis test, Wilcoxon rank-sum test with Bonferroni adjustment, and Spearman correlations were all performed as statistical and correlation analyses.

    Results

    The implant diameter significantly influenced pBICA and BICA, but not 3D BIC%. ITV and PTV were more sensitive to implant diameter than was ISQ. The coefficients of determination were high (>0.92) for the correlations between pBICA (or BICA) and indexes of the primary implant stability.

    Conclusion

    This study revealed how the implant diameter and the three-dimensional (3D) BIC influence the primary stabilities of dental implant. ITV and PTV were more sensitively influenced by the implant diameter than ISQ. The pBICA and BICA seem to be more important than 3D BIC % for using wider implant in treatment plan, since those two parameters are highly predictive of variations in the primary stability of dental implant.
    關聯: JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION
    顯示於類別:[生物資訊與醫學工程學系 ] 期刊論文

    文件中的檔案:

    檔案 大小格式瀏覽次數
    index.html0KbHTML232檢視/開啟


    在ASIAIR中所有的資料項目都受到原著作權保護.


    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回饋